$1,528.00
per spectrometer
| Configuration | |
|---|---|
| Model | ATP2000P |
| Spectral range | 200 nm-1000 nm |
| Slit width | 25 micron |
| Detector | UV-enhanced linear CMOS, 2048 effective pixels |
| Optical path | |
| Design | F/4 crossed asymmetric Czerny-Turner |
| Input | SMA905 fiber interface |
| Resolution | 0.1 nm-3 nm, dependent on slit and spectral range |
| Performance | |
| SNR | >450:1 in the current ATP2000-series datasheet |
| Dynamic range | >3000:1 |
| Integration time | 1 ms-65 s |
| Frame rate | Up to 300 fps |
| Electrical | |
| Data output | USB 2.0 or UART |
| ADC | 16 bit |
| Power | USB power supply, DC 5 V ±5% |
| Expansion interface | 20-pin dual-row programmable expansion interface |
| Operating environment | |
| Operating temperature | -10 °C to +40 °C |
| Humidity | <90% RH, non-condensing |
The ATP2000P is a higher-SNR UV-VIS spectrometer for measurements that benefit from more detector pixels while retaining a compact modular format. This configuration covers 200 nm-1000 nm with a 25 micron slit, a UV-enhanced 2048-pixel CMOS detector, and an F/4 crossed asymmetric Czerny-Turner optical path.
It is a strong fit for absorbance, transmission, fluorescence, LED characterization, fiber-laser LIBS, and general-purpose spectroscopy setups where broad UV-visible coverage and straightforward USB/UART integration matter.
If you need slit widths or wavelength ranges not available in the listed configuration, please contact us at [email protected] for a custom quote.